Agilent Technologies 856290216 Water Dispenser User Manual


 
583 RBW 30K
Unable to adjust 30 kHz resolution bandwidth.
584 RBW
IOOK
Unable to adjust 100 kHz resolution bandwidth.
585 RBW 300K
Unable to adjust 300 kHz resolution bandwidth.
586 RBW
IM
Unable to adjust 1 MHz resolution bandwidth.
587 RBW 30K
Unable to adjust 30 kHz resolution bandwidth.
588 RBW
IOOK
Unable to adjust 100 kHz resolution bandwidth.
589 RBW 300K
Unable to adjust 300 kHz resolution bandwidth.
590 RBW
IM
Unable to adjust 1 MHz resolution bandwidth.
591 LOG AMPL
Unable to adjust amplitude of log scale.
592 LOG AMPL
Unable to adjust amplitude of log scale.
593 LOG TUNE
Limiter calibration tune error from DC logger calibration.
594 LOG OFST
Attenuator calibration offset error from DC logger calibration.
595 LOG ATTN
596 LOG FID
597 LOG OFST
598 LOG OFST
599 LOG GAIN
Attenuator calibration absolute error from DC logger calibration.
Fidelity error from DC logger calibration.
Fidelity offset error from DC logger calibration.
Fidelity offset unstable from DC logger calibration.
Fidelity gain error from DC logger calibration.
System
Errors
(600
to
651)
600 SYSTEM
601 SYSTEM
ADC timeout errors occur if the A2 controller assembly frequency
counter is faulty. Refer to Chapter 8, “ADC/Interface Section.”
Hardware/firmware interaction; check other errors.
Hardware/firmware interaction; check other errors.
650
OUTOF
RG
651 NO
IRQ
ADC input is outside of the ADC range.
Microprocessor is not receiving interrupt from ADC.
Digital
and
Checksum
Errors
(700
to
799)
EEROM Checksum Errors (700 to 704)
Faults on the A2 controller assembly can cause these errors. Refer
to Chapter 10, “Controller Section.” Although some of these errors
might result in a blanked display, it is possible to read these errors
over HP-IB. Refer to “Troubleshooting to a Functional Section” in this
chapter.
The EEROM on A2 is used to store data for frequency response
correction, elapsed time, focus, and intensity levels. Error codes from
700 to 703 indicate that some part of the data in EEROM is invalid.
An EEROM error could result from either a defective EEROM or an
General Troubleshooting 7-37