Table 1-4. Recommended Test Equipment (continued)
Instrument
Counters
Frequency standard
Critical Specifications
for Equipment Substitution
Output frequency: 10 MHz
Accuracy:
<l
x
10-l’
Recommended
Use
Model
HP 5061B
PA
Microwave frequency counter
Universal counter
Frequency range: 9 MHz to 7 GHz HP 5343A OR
PAT,
External frequency reference input HP 5350B*
WV
Timebase accuracy (aging):
<5
x
lo-“/day
Option 001
Modes: TI
A+B,
frequency count HP
5334A/B
P
Time interval measurement range: 100 ns to 120 s
Frequency count range: 400 Hz to 11 MHz
Frequency resolution: 1 mHz
Timebase accuracy (aging):
<3
x
10m7/month
External 10 MHz reference input
Receivers
spectrum
analyzer Frequency range: 300 kHz to 7 GHz
Amplitude range: -70 dBm to
+20
dBm
HP 8566B*
A,T
VIeasuring
receiver
Compatible w/power sensors
dB relative mode
Resolution: 0.01 dB
Reference accuracy:
<fl.2%
HP
8902A*
PAT,
WV
iensors
‘ower
sensor Frequency range: 10 MHz to 13.2 GHz
Maximum SWR:
1.40 (10 to 30 MHz)
1.18 (30 to 50 MHz)
1.10 (50 MHz to 2 GHz)
1.18 (2 to 13.2 GHz)
HP 8481A+
PAT,
WV
‘ower sensor
Frequency range: 250 MHz to 350 MHz
Power range: 100
nW
to 10
PW
Maximum SWR: 1.15 (250 to 350 MHz)
Frequency range: 100 kHz to 2.9 GHz
Maximum S WR:
1.1 (1 MHz to 2.0 GHz)
1.30 (2.0 GHz to 2.9 GHz)
HP 8481D
P,A
‘ower sensor
HP
8482A*
PAT,
WV
Part of microwave workstation
=
performance tests; A
=
adjustments; M
=
test
&
adjustment module; T
=
troubleshooting;
J
=
operation verification
General Information l-l 1