2715 Spectrum Analyzer Service Manual
Use the following procedure to adjust R640 and R740 on the Sweep board.
H Test equipment required:
TV V isual Carrier Source
Oscilloscope with Probes
Multimeter (DMM)
1. Press UTIL 0 to initialize instrument settings. Set SPAN/DIV to 2 MHz and
RESOLUTION BW to 5 MHz.
2. Connect an RF visual carrier (such as from a TV antenna) to the RF INPUT.
3. Tune the Spectrum Analyzer to a TV visual carrier via UTIL 2 0. See Table
5--5 for visual carrier frequencies.
Table 5- 5: Television Channels
Channel Video Carrier
2 55.25 MHz
3 61.25 MHz
4 67.25 MHz
5 77.25 MHz
6 83.25 MHz
7 175.25 MHz
8 181.25 MHz
9 187.25 MHz
10 193.25 MHz
11 199.25 MHz
12 205.25 MHz
13 211.25 MHz
NOTE. The aural carrier is 4.5 MHz away from the visual carrier. Carrier
frequencies for Channels A through W and 14 through 83 are omitted from
this table.
4. Set video signal parameters by pressing DEMOD 9 and selecting BROAD-
CAST VIDEO DETEC T MODE, POSITIVE SYNC POLARITY, and
NEGATIVE VIDEO POLARITY. Press any menu button to exit the menu.
5. Set the peak carrier level at the top graticule line (REF LEVEL or ).
6. Enable the broadcast (AM) video mode by pressing DE MOD 3, and set the
LEVEL control for minimum flicker in the displayed picture.
Video Monitor Adjustment