Adjustment Procedures
2715 Spectrum Analyzer Service Manual
5-43
o. Set the peak carrier level at the top graticule line.
p. Enable the FM video mode by pressing DEMOD 3, and set the LEVEL
control for minimum flicker in the displayed picture.
q. Adjust R740 on the Sweep board for best contrast in the picture.
r. Initialize instrument settings by pressing UTIL 1 1.Ifuserdefined
power up settings exist, the instrument will recall those settings. If no
user defined power up settings exist, the instrument will recall factory
default power up settings.