Maintenance
2715 Spectrum Analyzer Service Manual
6-3
deionized water. Abrasive cleaners should not be used. If the circuit board
assemblies need cleaning, remove the circuit board by referring to the instruc-
tions under Corrective Maintenance on Page 6--13.
After cleaning, allow the interior to dry thoroughly before applying power to the
instrument.
NOTE. Do not allow water to get inside any enclosed assembly or components
such as the hybrid assemblies, RF Attenuator assembly, potentiometers, etc.
Instructions for removing these assemblies are provided in the Corrective
Maintenance on Page 6--13 part of this section. Do not clean any plastic
materials with organic cleaning solvents such as benzene, toluene, xylene,
acetone, or similar compounds because they may damage the plastic.
The following is a list of tools and fixtures that are needed to service the
Spectrum Analyzer:
H Screwdriver, flat, with 1/4 to 3/8 inch bit
H Screwdriver, Posidriv 440-2
H Screwdrivers, Torx T15 and T20
H Wrench, 5/16 inch open end
H Hex drive wrenches, 3/32, 5/64,7/64 inch
H Torque W rench Kit
An optional service kit (the Extender Kit) is available to aid in servicing the
Spectrum Analyzer. The Extender Kit consists of the following:
H Main Extender Board
H Small Extender Board
H Extender Board for the 300 Hz Bandpass Filter board
H Three SMB male to female cables
H Two SMA male to male cables
H Two SMA female to female adapters
H Female BNC to female SMB cable
H Male BNC to square pin cable
H Data sheet
Fixtures and Tools for
Maintenance