Tektronix 2715 Water Dispenser User Manual


 
Adjustment Procedures
2715 Spectrum Analyzer Service Manual
5-53
For this adjustment the 2715 is controlled by a personal computer using the
RS-232 or GPIB interface. The test equipment is controlled by GPIB. If
necessary, refer to the documentation provided with the GPIB interface card for
installation instructions.
The following steps describe the installation process.
H Install the test software
H Run the system configuration program (CONFIG)
H Connect the personal computer to control the test instrument
H Run the test equipment characterization program (CHAR)
NOTE. These steps are required for initial installation only. After the initial
installation, it is not necessary to run CONFIG or CHAR as long as the system
remains intact.
Install the Flatness Test Software. The Flatness Test software is contained on a set
of floppy disks (271X Flatness Tests) supplied with this manual. Use the
following procedure to install the software onto the personal computer:
1. Install disk 1 in drive A of the personal computer.
2. Set the default drive to A by typing:
A:<Return>
3. Type the following command to install the flatness software on your hard
disk:
Install<Return>
A prompt will instruct you when to switch disks (for 5.25 disks only).
4. Reboot your P C to incorporate the changes to the autoexec.bat file.
NOTE. INSTALL.BAT is a batch file that creates a directory on your hard disk
(called TEKCATS) and several subdirectories. It then copies the Flatness Test
software into these directories.
Computer Setup and
Software Installat ion