9- 187
2715 Spectrum Analyzer Service Manual
2715 settings are:
100 MHz Center Frequency
2 MHz span
5 dB/div
INPUT MENU 1 (cal signal on)
DSPL MENU 4 (peak mode)
Readouts off
D waveform on
Intensity on
Adjust signal to top of screen with ref level and ref level fine.
TP924
Schematic location: 3B4
DSHOR
Schematic location: 4B1
TP418
Schematic location: 5C3
UNBLAN
K
Level
BLANK
Level
TP120
Schematic location: 5A3
TP250
Schematic location: 2B2
TP270
Schematic location: 1C5
DSVER
Schematic location: 3C2
UNBLANK
Level
B
L
A
N
K
Figure 1-41: A15A1 Power Supply Waveforms