Tektronix 2715 Water Dispenser User Manual


 
Specifications
1-22
2715 Spectrum Analyzer Service Manual
Table 1- 9: Environmental Characteristics (Cont.)
Characteristic Description
Vibration
Operating
(Instrument secured to a vibration
platform during test)
MIL-T-28800E, Method 514 Procedure X (modi fied) 15 minutes along each of 3
major axes at a total displacement of 0.015 in (0.38 mm) peak-to-peak (2.4 g at
55 Hz), with frequency varied from 10 Hz to 55 Hz in 1-minute sweeps
Hold for 10 minutes at 55 Hz
All major resonances must be above 55 Hz (resonance is defined as an excursion
greater than 2X t he input displacement)
Shock (Operating and Nonoperating) Three guillotine-type shocks of 30 g, one-half sine, 11 ms duration each direction
along each major axis, total of 18 shocks; no drops allowed on the front surface and
front corners
Transi t Drop (Free Fall) 8 in (203.2 mm), one per each of 5 faces and 4 corners (instrument is tested and
meets drop height of 12 i n (304.8 mm)
Table 1- 10: Physical Charact eristics
Characteristic Performance Requirement
Weight
With Standard Accessories 25 lbs (11.34 kg)
Without Standard Accessories 22.5 lbs (10.21 kg)
Dimensions
Height with Feet and Handle 5.4 in (137.16 mm)
Width
With Handle 14.2 in (360.68 mm)
Without Handle 12.9 in (327.66 mm)
Depth
With Front Panel Cover 17.5 in (444.50 mm)
Without Front Panel Cover 16.85 in (427.99 mm)
With Handle Extended 20.1 in (510.54 mm)