Operating Examples
AWG2021 User Manual
2 Ć79
2. Select Frequency from the side menu.
3. Press 1 and the MHz/ms key in the Unit key to input the frequency with the
numeric keys.
4. To set the frequency with the general purpose knob, select the index digit for
input with the front panel arrow buttons (/). Press the button to move
the underscore to the left or press the button to move the underscore to
the right. Turn the general purpose knob with the index digit selected to get a
1.000 MHz frequency.
In the same manner as the frequency, above, you will set the amplitude to 5 V
and the offset to 0 V.
5. Select Amplitude from the side menu.
Use the general purpose knob or numeric keys to set an amplitude of
5.000 V.
6. Select Offset from the side menu.
Turn the general purpose knob and check that the 0 V line (broken line) on
the screen moves within a range ±2.5 V. Then set the offset to 0.000 V. You
could also input the offset value with the numeric keys.
7. Check that when you press the Polarity button in the side menu, the polarity
toggles between Normal and Invert and the polarity of the sine wave on the
screen is reversed. Then set the polarity to Normal.
Figure 2 -67 shows the screen when the output parameters are set for a sine
wave on CH1.