Appendix C: Performance Verification
AWG2021 User Manual
CĆ11
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Select the file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to display the list of waveform files
and highlight the MODE.WFM file.
H Push ENTER to select the file. This button is located to the
lower-right of the numeric keypad.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check cont mode: Check that the amplitude of the sine wave displayed on
the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are
displayed.
5. If Option 02 is installed (Option 02 adds the CH2 output channel): Repeat
this procedure, connecting the oscilloscope to the AWG2021 CH2 output
connector, pushing Waveform Sequence to select the waveform for CH2,
and turning on the CH2 output.
6. End procedure: Disconnect the oscilloscope.
Electrical
Characteristic Checked
Operating mode, Triggered, on pageĂBĆ1.
Equipment Required Two 50ĂW coaxial cables, a function generator, and an
oscilloscope.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Hook up the function generator:
H Connect the AWG2021 TRIGGER INPUT connector through a
coaxial cable to the function generator output connector (see
Figure C-4).
Check Triggered Mode