LOAD/SAVE Menu
AWG2021 User Manual
3 Ć249
Selecting the Device
Select Device to select the source from which files are loaded into the AWG2021
internal memory and the destination to which files are saved from internal
memory. You may select Disk, NVRam or GPIB.
1. Select Device from the bottom menu.
2. Select Disk, NVRam or GPIB from the side menu.
The following items are listed in the side menu:
Disk A floppy disk. Files are saved to or loaded from a floppy
disk inserted into the floppy disk drive on the right side of
the instrument.
NVRam The instrument’s internal non-volatile RAM.
GPIB The GPIB interface. Used to transfer waveform data directly
from another instrument to this instrument through a GPIB
cable.
To transfer waveform data through the GPIB interface, the remote port must be
set to GPIB and the GPIB operating mode must be set to Waveform Transfer.
If this is not done, the following message will appear when you select the GPIB
item:
The GPIB configuration is not Waveform Transfer"
and the Remote Port is not GPIB".
Are you sure of changing the parameters
to match the transfer operation ?
Selecting O.K. in the sub-menu at this point will cause these items to be set to
the above items automatically.
Table 3-19 shows a list of instruments for which direct transfer of waveform data
to this instrument is supported.