with the exception of Date Time (date and time) to be restored to their default
settings. Default settings consist of the factory settings and the following items:
Remote Port GPIB
GPIB Operating Mode Talk/Listen
GPIB Address 1
RS-232-C Parameters
Baudrate 9600
Data Bits 8
Parity None
Stop Bits 1
Flagging None
CAUTION. Once deleted, data cannot be restored.
1. Select Misc from the bottom menu.
2. Select Config... from the side menu.
3. Select Secure Erase Memory from the sub-menu.
4. The following message will appear, asking you to confirm that you really
want to delete the data from memory:
Initializes both catalog memory and NVRam
then puts all setups in the factory defaults.
Answer Cancel or O.K. If you select O.K., all data will be deleted from
memory and settings will be restored to factory defaults, and then the SETUP
menu will appear.
Remote Port Settings. The instrument’s rear panel has two remote control
interface ports: IEEE STD 488 (GPIB) and RS-232-C. (Note: There is no
RS-232-C port when Option 03 or Option 04 has been installed.) Select the
desired port, depending on which interface you will be using.
1. Select Misc from the bottom menu.
2. Select Config... from the side menu.
3. Press the Remote Port button in the sub-menu and select either GPIB or
RS232C. The remote interface port that you have selected will be displayed
on the left side of the status line at the top of the screen.