Appendix C: Performance Verification
CĆ10
AWG2021 User Manual
Operating Mode Checks
These procedures check operation of the Cont, Triggered, Burst, Gated,
Waveform Advance, and Autostep modes.
Electrical
Characteristic Checked
Operating mode, Continuous, on pageĂBĆ1.
Equipment Required A 50ĂW coaxial cable and an oscilloscope.
Prerequisites The AWG2021 must meet the prerequisites listed on
page CĆ5.
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure C-3).
OscilloscopeAWG2021
AWG2021
540
TDS
Figure CĆ3: Cont Mode Initial Test Hookup
b. Set the oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling: DC. . . . . . . . . . . . . . . .
CH1 scale 0.2 V/div.. . . . . . . . . . . . . . . . . . . .
CH1 input impedance: 50 W. . . . . . . . . .
Horizontal
Sweep 500 ms/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level +100 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Auto. . . . . . . . . . . . . . . . . . . . . . .
Check Cont Mode