AWG2021 User Manual
3 Ć181
H Clock Settings. Select
using the general purpose knob.
Then press the VALUE button on the front panel. The menu shown in Figure
3 -131 will appear.
Figure 3 Ć131: Clock Setting Menu
Set the clock source and the clock frequency by selecting the appropriate items
in the side menu. The settings for clock source and clock frequency are used for
both channel 1 and channel 2.
Internal Clock Selects the internal clock. Use the numeric keys or
the general purpose knob to set the internal clock
frequency.
External Clock Selects the external clock. The external clock signal
is input through the CLOCK IN connector on the
rear panel of the instrument.
Default Value Sets the internal clock frequency to the default value
of 100.0 MHz.
After setting the clock source and clock frequency,
select O.K. from the side menu. The values for