Tekxon Technology AWG2021 Ventilation Hood User Manual


 
EDIT Menu
AWG2021 User Manual
3 Ć91
As there is less data in the Source waveform than the destination waveform, data
at level 0 (in other words, a straight line) has been automatically added from
point 626 to point 875 in the Source waveform.
0 125 625 875
Destination Waveform
Comparison Area
Source Waveform
Comparison Area
Figure 3 Ć70: Comparison Without Hysteresis
Comparison With Hysteresis — The left part of Figure 3 -71 shows a
hysteresis comparison between a triangular wave as Waveform2 and a square
wave as Waveform1, with the results output to the MARKER1 display area in
the Waveform2 area. The figure on the right has been provided as an aid to
understanding this process.
In hysteresis comparison, the waveform only becomes high level when the
destination waveform exceeds the level above the Source waveform set for
Hysteresis. The waveform only becomes low level when the destination
waveform exceeds the level below the Source waveform set for Hysteresis.
Comparison is only performed for the section of the waveform between the
vertical bar cursors (in this example, from point 0 to point 999).