4. Select Execute from the side menu. The selected diagnostics items are
executed.
If the test finishes without a problem, Pass is displayed on the Result
column. If an error occurs, Fail is displayed. If the instrument fails a test, an
error code is displayed in the Code column.
Calibrations. This instrument is equipped with the system to calibrate itself. This
enables the AWG2021 to operate with greater precision. A series of calibrations
is carried out by selecting the Calibrations item.
NOTE. The AWG2021 must complete its warm up (about 20 minutes) and
stabilize in a 20_ C to 30_ C environment before calibration. When the
instrument is powered off while the calibrations is executed, the calibration data
in the memory may be loss.
When the Calibrations item is selected, the list of calibration items shown in
Figure 3 -198 will appear.
Figure 3 Ć198: Calibration List
The calibration menu is divided into three columns: Calibrations, Result, and
Code.
The calibration items are executed individually or all together.
Setup CH2 is valid when Option 02 is installed. The Result and Code columns
are the same as for the diagnostics menu.
1. Select Diag/Cal from the bottom menu.