Tekxon Technology AWG2021 Ventilation Hood User Manual


 
AWG2021 User Manual
1 Ć3
Start Up
This section describes the procedures required prior to turning on the AWG2021.
Installation
Before you begin, refer to the Safety Summary at the front of this manual for
power source, grounding, and other safety information.
Before you use the instrument, ensure that it is properly installed and powered
on. To properly install and power on the instrument, perform the following steps:
1. Check that the operating environment is correct.
The AWG2021 operates correctly in ambient temperatures from +10_ C to
+40_ C and relative humidity from 20% to 80%. If this instrument is stored
at temperatures outside this usage temperature range, do not switch on the
power until the chassis has come within the usage temperature range. For the
information on the other operating environment, see Appendix B: Perfor-
mance Characteristics.
NOTE. If you are installing this instrument in the dedicated rack, refer to the
instruction sheet that comes with the rack mounting kit.
2. Before switching on the power, double check that there is nothing blocking
the flow of air at the fan and air intake holes.
This instrument takes in outside air and cools itself by forcibly exhausting
air with the fan on its left side. Leave space at the sides of this instrument so
that the heat generated within the instrument does not build up and harm the
operation. There are holes for air intake on the sides and bottom of this
cabinet. After switching on the power, double check that the fan is turning.
Here are the minimums for the space at the sides of this instrument.
Top and rear: 7.6cm (3 inches)
Left and right: 15.2cm (6 inches)
WARNING. Always unplug the power cord from the socket before checking the
line fuse to avoid electrical shock.