Appendix C: Performance Verification
AWG2021 User Manual
CĆ37
b. Set oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 Coupling DC. . . . . . . . . . . . . . . .
CH1 Scale 200 mV/div.. . . . . . . . . . . . . . . . . . .
CH1 Input Impedance 50 W. . . . . . . . . .
Horizontal
Sweep 50 ns/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level 500 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Auto. . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Select waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to highlight the SYNC_MRK.WFM
file.
H Push ENTER.
3. Check front-panel SYNC Out and MARKER Out amplitude:
a. Check SYNC Out pulse amplitude: Check that the pulse amplitude of the
waveform displayed on the oscilloscope is 1.2 V
p-p
or greater.
b. Check CH1 MARKER 1 Out pulse amplitude:
H Move the coaxial cable from the AWG2021 CH1 SYNC Out
connector to the CH1 MARKER 1 connector.
H Change the oscilloscope sweep to 5 ms/div.
H Check that the pulse amplitude of the displayed waveform is from
2.250 V
p-p
to 2.625 V
p-p
.