Tekxon Technology AWG2021 Ventilation Hood User Manual


 
EDIT Menu
AWG2021 User Manual
3 Ć89
2. Select Waveform1 from the side menu. This waveform will be the reference
for comparison.
3. Press the CURSOR button on the front panel.
4. Using the numeric keys or the general purpose knob, set the positions of the
left and right vertical bar cursors to define the section of the waveform for
comparison. In this example, we will set the left cursor to point 125 and the
right cursor to point 625.
NOTE. The value indicated by D shows the difference in points between the left
and right vertical bar cursors. This value includes the point occupied by the
cursor, so the number of points in the waveform is D + 1.
5. Select Waveform2 from the side menu. This waveform will be compared to
the reference waveform.
6. Press the CURSOR button on the front panel.
7. Using the numeric keys or the general purpose knob, set the positions of the
left and right vertical bar cursors to define the section of the waveform for
comparison. In this example, we will set the left cursor to point 125 and the
right cursor to point 875.
8. Select Operation from the bottom menu.
9. Select Compare... from the fourth page of the side menu (More 4 of 4).
Figure 3 -69 shows the menu with the Compare... item selected.
When three waveform areas have been designated in the waveform editor,
select one of the remaining two editing areas as the source for waveform
comparison with Source item in the sub-menu.