SETUP Menu
3 Ć222
AWG2021 User Manual
Multiplier
Amplitude 5 V
Offset 0 V
CH1 Output
Parameter Settings
CH1 Waveform
-2.5V
2.5V
CH1 Output Waveform
CH2 Waveform
5V
0V
-5V
0V
Amplitude 5 V
Offset 2.5 V
CH2 Output
Parameter Settings
Figure 3 Ć156: Multiplying CH1 Waveform by CH2 Waveform
This multiplier has a bandwidth of 30 MHz for the carrier (signal) and a
bandwidth of 4 MHz for the modulation signal. The waveform output from CH1
is the result of the following calculation.
CH1out = CH1 setting value × (CH2 setting value + CH2 offset)
A CH2 signal of 5 V (full scale) gives 100% modulation. When the CH2 signal
is minus, the CH1 signal is inverted.
The multiplied value of CH2 is expressed as a % and is displayed at the bottom
right of the CRT screen (see next example).
Max:
Min:
100.00%
0.00%
NOTE. When multiplication results in an output that exceeds 5 Vp-p, the signal
may be distorted.
Add The CH1 waveform is added to the waveform output from
CH2. See Figure 3 -157.