Appendix C: Performance Verification
AWG2021 User Manual
CĆ27
3. End procedure:
a. Disable function generator output: Turn the function generator output
off.
b. Remove equipment: Disconnect connections to the test equipment.
Clock Frequency and Amplitude Checks
These procedures check the accuracy of the AWG2021 clock frequency and the
waveform output amplitude.
Electrical
Characteristic Checked
Clock Generator, Accuracy, on pageĂBĆ6.
Equipment Required A 50ĂW coaxial cable, an SMAĆBNC adapter, and a frequency
counter.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install test hookup and set test equipment controls:
a. Hook up frequency counter: Connect the AWG2021 rear panel CLOCK
OUT connector to the frequency counter input through a coaxial cable
and an SMA-BNC adapter (see Figure C-13).
AWG2021
AWG2021
Frequency Counter
DC5010
Figure CĆ13: Clock Frequency Accuracy Initial Test Hookup
Check Clock Frequency
Accuracy