Appendix C: Performance Verification
CĆ30
AWG2021 User Manual
b. Set oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Scale 500 mV/div.. . . . . . . . . . . . . . . . . . . . . . . .
Input impedance 50 W. . . . . . . . . . . . . . .
Horizontal
Sweep 500 ns/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level 500 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Auto. . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Select the waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to select the CLK_AMPL.WFM
file.
H Push ENTER.
3. Check clock amplitude accuracy: Check that the pulse displayed on the
oscilloscope has an amplitude of 0.7 V
p-p
to 1.3 V
p-p
.
4. End procedure: Disconnect the oscilloscope.