EDIT Menu
3 Ć58
AWG2021 User Manual
Operation (More 3 of 3) Marker...
Set Pattern
Set High
Set Low
Clear Pattern
Cancel
O.K.
Key Data
1 Bit
4 Bits
Import Line Data /
User defined
Code Config...
Initial Src
Initial Code
Out [I/O]
High/Low
Invert/Keep
0
0
1
1
Go Back
1
2
Marker
NOTE. When a new waveform file is created, the marker signal for the first point
of the waveform data is set to high as a default value.
The output level for the marker signals is 1.2 V with a termination of 50 W. The
marker signals for CH1 are output from the CH1 MARKER 1 output connector
on the front panel and CH1 MARKER 2 OUT connector on the rear panel,
respectively. Similarly, when Option 02 is installed, the CH2 marker signals are
output from the CH2 MARKER 1 OUT and CH2 MARKER 2 OUT
connectors, respectively, on the rear panel.
The marker level changes to low after waveform output is complete or when the
STOP button in the MODE menu is pressed.
1. Move the left and right vertical bar cursors to define the area for marker
setting, then select Marker... from the third page of the side menu (More 3
of 3).
2. Press the Marker button in the side menu to select marker 1 or marker 2.
3. Press the sub-menu Set High or Set Low button to set the desired marker
state. Figure 3 -35 shows marker signal set high between the vertical bar
cursors.
H Set High. The marker signal for the section of the waveform between the
vertical bar cursors will be set to High.
H Set Low. The marker signal for the section of the waveform between the
vertical bar cursors will be set to Low