EDIT Menu
3 Ć52
AWG2021 User Manual
AfterBefore
Figure 3 Ć29: Vertically Shifted Display
9. Select Go Back from the sub-menu.
Scaling Function. Use Scale... to change the scale for the waveform data between
the vertical bar cursors. Executing scaling changes the waveform data. The
following diagram shows the menu configuration for the Scale... item.
Operation (More 2 of 3) Scale...
Scale
Factor
New Size
Origin
Horizontal
Vertical
Go Back
Execute
1. Move the left and right vertical bar cursors to define the editing area for
scaling, then select Scale... from the second page of the side menu (More 2
of 3).
2. Press the Scale button in the sub-menu to select Horizontal.
The Scale sub-menu item selects the direction in which the data between the
vertical bar cursors is scaled.
3. Select Factor or New Size from the sub-menu.
The magnification for scaling can be set with either Factor or New Size.
H Factor. The waveform data between the vertical bar cursors is expanded/
reduced by this ratio. This parameter is linked with New Size. Factor can
be set within the range ±100 in steps of 0.01. However, the total number