Appendix C: Performance Verification
CĆ20
AWG2021 User Manual
H Push MODE!Waveform Advance!Slope to highlight Positive.
H Select Level from the side menu, and turn the general purpose knob
to select a 1.0 V level.
H Select Impedance from the side menu to highlight 50 W.
c. Select waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Highlight the MODE_ADV.SEQ file using the general purpose
knob.
H Push ENTER to select the file.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check waveform advance: Repeatedly push the AWG2021 MANUAL
TRIGGER button, and check that the oscilloscope displays a continuous
sine wave that switches between two frequencies at each manual trigger.
5. End procedure: Disconnect the oscilloscope.
Electrical
Characteristic Checked
Operating mode, Autostep, on pageĂBĆ1.
Equipment Required Two 50ĂW coaxial cables and an oscilloscope.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope:
H Connect the AWG2021 CH1 output through a coaxial cable to the
oscilloscope CH1 vertical input.
H Connect the AWG2021 CH1 SYNC Out output through a coaxial
cable to the oscilloscope CH1 vertical input (see Figure C-10).
Check Autostep Mode