Appendix C: Performance Verification
CĆ50
AWG2021 User Manual
Floating Point Processor Check (Option 09)
This procedure checks the AWG2021 floating point processor. This check
requires that the AWG2021 has Option 09 installed.
Equipment Required None.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Check that floating point processor test in internal diagnostics passes:
a. Run the AWG2021 internal diagnostics: Push the AWG2021 ON/STBY
switch two times so that the AWG2021 runs the power-on diagnostics.
b. Check the FPP test results: When the AWG2021 finishes the FPP test,
check that the test result is Pass.
This completes the performance tests for the AWG2021.