Tekxon Technology AWG2021 Ventilation Hood User Manual


 
Appendix C: Performance Verification
AWG2021 User Manual
CĆ9
Table CĆ2: File List for Performance Check/Adjustment Disk (Cont.)
No. Usage
SETUP MenuEDIT Menu
File NameNo. UsageOffset
Amp
FilterOperationClock
Wfm
Point
Wfm
Shape
File Name
7 GAIN_OFF.AST
Step: 1 GAINĆ1.WFM
1000 1 MHz Normal Through 0.5 V 0V Gain
Accuracy
Step: 2 GAINĆ2.WFM 1000 1 MHz Normal Through 0.5 V 0V
Step: 3 GAINĆ3.WFM 1000 1 MHz Normal Through 5.0 V 0V
Step: 4 GAINĆ4.WFM 1000 1 MHz Normal Through 5.0 V 0V
Step: 5
OFFSETĆ1.WFM
1000 1 MHz Normal Through 0.05 V 2.5 V Offset
Accuracy
Step: 6
OFFSETĆ2.WFM
1000 1 MHz Normal Through 0.05 V 0V
Step: 7
OFFSETĆ3.WFM
1000 1 MHz Normal Through 0.05 V -2.5 V
8 PULSE.WFM 64 250 MHz ĊĊ Through 0.5 V 0V Pulse
Response
9 SYNC_MKR.WFM 200 1 MHz ĊĊ Through 1V 0V SYNC Out and
MARKER OUT
Amplitude
10 TRG_IN.WFM 1000 100 MHz ĊĊ Through 1V 0V External Trigger
Level Accuracy
11 EXT_CLK.WFM 1000 External
Clock
ĊĊ Through 1V 0V External Clock
In
12 DIGI_OUT.WFM 4096 1 MHz Normal Through 1V 0V ECL and TTL
DIGITAL DATA
OUT Check
(Option 03 and
04)