Appendix C: Performance Verification
AWG2021 User Manual
CĆ17
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Modify the AWG2021 default settings:
H Push MODE!Gated!Polarity to highlight Positive.
H Select Impedance from the side menu to highlight 1 MW.
c. Select the file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Highlight the MODE.WFM file, using the general purpose knob.
H Push ENTER to select the file.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check gated mode with manual trigger: Push and hold the AWG2021
MANUAL TRIGGER button, and check that the oscilloscope continuously
displays a sine wave while the MANUAL TRIGGER button is pushed.
5. Check gated mode with gate signal:
a. Apply gate signal: Turn function generator output on.
b. Check gated mode with positive gate signal: Check that the oscilloscope
displays a sinewave when the gate signal is high.