Appendix C: Performance Verification
AWG2021 User Manual
CĆ23
Function Generator
AWG2021
FG5010
DMM
8842A
AWG2021
Figure CĆ11: External AM Operation Initial Test Hookup
c. Set DMM controls:
Mode VDC. . . . . . . . . . . . . . . . . . . . . . . . . . .
Range 20. . . . . . . . . . . . . . . . . . . . . . . . . .
Inputs Front. . . . . . . . . . . . . . . . . . . . . . . . . .
d. Set function generator controls:
Function Square. . . . . . . . . . . . . . . . . . . . . . . .
Mode Continuous. . . . . . . . . . . . . . . . . . . . . . . . . . .
Parameter
Frequency 1 kHz. . . . . . . . . . . . . . . . . . . .
Amplitude 0 V. . . . . . . . . . . . . . . . . . .
Offset 1 V. . . . . . . . . . . . . . . . . . . . . . .
Output Off. . . . . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Modify AWG2021 default settings:
H Push MODE!Autostep.
c. Select waveform file:
H Select Select Autostep File from the side menu to choose a
waveform file for CH1.