Tekxon Technology AWG2021 Ventilation Hood User Manual


 
Appendix C: Performance Verification
AWG2021 User Manual
CĆ31
Gain and Offset Accuracy Checks
These procedures check the accuracy of the AWG2021 gain and offset.
NOTE. The gain and offset accuracy checks are structured as a continuous test.
After Check Gain Accuracy, the next test uses the control settings from the last
test and uses the next step in the sequence file.
Electrical
Characteristic Checked
Main Output, Amplitude, Accuracy, on pageĂBĆ6.
Equipment Required A 50ĂW coaxial cable, a BNCĆtoĆdual banana adapter, and a
digital multimeter (DMM).
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install test hookup and set controls:
a. Hook up DMM: Connect the AWG2021 CH1 output through a 50 W
coaxial cable, a 50 W terminator and a dual banana connector to the
DMM INPUT connector (see Figure C-15).
DMM
AWG2021
8842A
AWG2021
Figure CĆ15: Gain Accuracy Initial Test Hookup
b. Set DMM controls:
Mode VDC. . . . . . . . . . . . . . . . . . . . . . . . . . .
Range 20. . . . . . . . . . . . . . . . . . . . . . . . . .
Input Front. . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
Check Gain Accuracy