Tekxon Technology AWG2021 Ventilation Hood User Manual


 
Appendix C: Performance Verification
AWG2021 User Manual
CĆ19
Electrical
Characteristic Checked
Operating mode, Waveform Advance, on pageĂBĆ1.
Equipment Required A 50ĂW coaxial cable and an oscilloscope.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope (see Figure C-9).
OscilloscopeAWG2021
AWG2021
540
TDS
Figure CĆ9: Waveform Advance Mode Initial Test Hookup
b. Set oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling DC. . . . . . . . . . . . . . . . .
CH1 0.2 V/div.. . . . . . . . . . . . . . . . . . . . . . . .
CH1 input impedance 50 W. . . . . . . . . .
Horizontal . . . . . . . . . . . . . . . . . . . . . . . .
Sweep 5 ms/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level 0 V. . . . . . . . . . . . . . . . . . . . . . .
Mode Auto. . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Set AWG2021 controls:
Check Waveform Advance
Mode