Appendix C: Performance Verification
CĆ14
AWG2021 User Manual
Electrical
Characteristic Checked
Operating mode, Burst, on pageĂBĆ1.
Equipment Required A 50ĂW coaxial cable and an oscilloscope.
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure C-5).
OscilloscopeAWG2021
AWG2021
540
TDS
Figure CĆ5: Burst Mode Initial Test Hookup
b. Set the oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling: DC. . . . . . . . . . . . . . . .
CH1 scale 0.2 V/div.. . . . . . . . . . . . . . . . . . . .
CH1 input impedance: 50 W. . . . . . . . . .
Horizontal
Sweep 500 ms/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level +100 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Norm. . . . . . . . . . . . . . . . . . . . . . .
2. Set AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Modify the AWG2021 default settings: Push MODE!Burst!Burst
Count and turn the general purpose knob to a burst count of 3.
Check Burst Mode