Appendix C: Performance Verification
CĆ6
AWG2021 User Manual
memory. For instructions on loading files, see Loading Sample Waveforms
on page 2 -28 in section 2.
Read Preparation and Conventions on page C-1. Also, if you are not familiar
with operating the AWG2021, read in section 2 before doing any of these
procedures.
The following equipment is required to check the performance of the AWG2021.
Table CĆ1: Test Equipment
Item Minimum Requirements Example Purpose
Precision termination Impedance: 50ĂW, 0.1%
Connectors: BNC
Tektronix Part 011Ć0129ĆXX Signal termination.
Adapter Connectors: BNC femaleĆtoĆdual
banana
Tektronix Part 103Ć0090ĆXX Signal interconnection.
Adapter Connectors: SMA maleĆtoĆBNC
female
Tektronix Part 015Ć0554ĆXX Signal interconnection.
BNC dual input (TEE) adapter Connectors: BNC Tektronix Part 103Ć0030ĆXX Signal interconnection.
BNC cable (4 required) Impedance 50ĂW
Connectors: BNC
Length: 43 inches
Tektronix Part 012Ć0057ĆXX Signal interconnection.
Termination board (Option 03) Must use example equipment Tektronix Part 671Ć2957ĆXX Used to check ECL digital
data output.
ECL Digital Data Out Cable
(Option 03)
Must use example equipment Tektronix Part 012Ć1408ĆXX
(Standard accessory for
Option 03)
Used to check ECL digital
data output.
Test leads (Option 03) Must use example equipment Tektronix Part 012Ć1381ĆXX Used to check ECL digital
data output.
TTL Digital Data Out Cable
(Option 04)
Must use example equipment Tektronix Part 174Ć3192ĆXX
(Standard accessory for
Option 04)
Used to check TTL digital
data output.
2 × 13 header (Option 04) Must use example equipment Tektronix Part 131Ć3847ĆXX Used to check TTL digital
data output.
Probe, 10X (Option 04) 10X probe Tektronix Part P6139A Used to check TTL digital
data output.
Test oscilloscope Bandwidth: >250ĂMHz Tektronix TDS500 Series
Digitizing Oscilloscope or
2400 Series Digitizing OscilloĆ
scope
Checks output signals. Used
in many procedures.
Frequency counter Frequency range: 10ĂHz to 250ĂMHz Tektronix DC 5010 ProgramĆ
mable Universal Counter/TimĆ
er
*
Used to check clock frequenĆ
cy.
Related Information
Equipment Required