Appendix C: Performance Verification
CĆ24
AWG2021 User Manual
H Turn the general purpose knob to highlight the OPE.AST file.
H Push ENTER.
3. Turn on the AWG2021 CH1 output: If the LED above the CH1 output
connector is not on, push the CH1 button.
4. Enable the function generator output: Turn on the function generator output.
5. Check external AM operation:
H Check that the step number displayed on the AWG2021 MODE menu is
Step 1 (see Figure C-12). If it is not, push MANUAL TRIGGER to step
though the sequence file steps until Step: 1 is displayed.
Step Display
Figure CĆ12: MODE Menu Autostep Setting
H Check that the DMM reading is in the range from 2.375 to 2.625 V
(100% modulation).
H Set the function generator offset value to 0 V. Check that the DMM
reading is in the range from 1.125 to 1.375 V (50% modulation).