Appendix C: Performance Verification
CĆ26
AWG2021 User Manual
Electrical
Characteristic Checked
Arithmetic Operation, Amplitude Modulation, on pageĂBĆ9.
Equipment Required Two 50ĂW coaxial cables, a function generator, and a digital
multimeter (DMM).
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5. It
also requires Option 02 (CH2 output channel) in the AWG2021.
1. Use test hookup and control settings from previous check.
2. Check internal AM operation:
a. Check Autostep Step 2:
H Push the AWG2021 MANUAL TRIGGER button, and check that
the step changes to Step 2 on the MODE menu.
H Check that the DMM reading is in the range from 2.375 VDC to
2.625 VDC.
b. Check Autostep Step 3:
H Push the AWG2021 MANUAL TRIGGER button, and check that
the Autostep changes to Step: 3 on the MODE menu.
H Check that the DMM reading is in the range from –2.625 V to
–2.375 V.
3. End procedure: Retain the test hookup and settings for the next check.
Electrical
Characteristic Checked
Arithmetic Operation, Add, on pageĂBĆ9.
Equipment Required Two 50ĂW coaxial cables, a function generator, and a digital
multimeter (DMM).
Prerequisites The AWG2021 meets the prerequisites listed on page CĆ5. It
also requires Option 02 (CH2 output channel) in the AWG2021.
1. Use test hookup and control settings from previous check.
2. Check internal Add operation:
a. Check Autostep Step 4:
H Push the AWG2021 MANUAL TRIGGER button, and check that
the step changes to Step 4 on the MODE menu.
H Check that the DMM reading is in the range from 2.85 V to 3.15 V.
Check Internal AM
Operation (Option 02 only)
Check Internal Add
Operation (Option 02 only)